img

Handling of Out of Calibration Equipments and Instruments

All instruments and equipment’s should be calibrated, verified and validated regularly. In case of any discrepancies where they are out of service then procedures and considerations need to be done.

Sharing is caring, show love and share the thread with your friends.

Description

Handling Out of calibration instruments

  • When any instrument is found to be out of tolerance then it should be immediately notified to the supervisor or the person responsible
  • The calibration data is to be reviewed to verify the validity of out of tolerance
  • Investigation is to be done on the root cause of out of calibration instruments
  • Historical data can be reviewed which can provide the factors leading to the root cause
  • Corrective and preventive action should be initiated regarding the instrument in terms of replacement or re calibration or repair or removal from operation
  • Notification regarding out of calibration should be given to owner or supervisor responsible for the instrument
  • The notification should include current calibration data, range of error and date of last calibration which was successful
  • The supervisor evaluates the root cause and its effect on quality of the product
  • The effect on quality of the product is evaluated based on the standard operating procedure which defines the quality standards
  • When initiating the CAPA supervisor should consider the process requirements which are measured by the instrument
  • Major decisions which are taken regarding its effect on quality should be approved by quality department and also be documented
  • The subsequent events should be observed. The complete process should be documented which can be useful in the future when similar problem arises.
  • Change control can be initiated which may require re execution or re calibration of instruments installation, operational and performance qualification
  • Proper review should be done with approval of quality departments

Handling Out of calibration Equipment’s

  • When equipment’s are mishandled or overloaded then they show defective results or which fall out of specified limits.
  • The equipment which is out of specified limits is isolated and labeled to avoid using it until repaired properly
  • Root cause analysis is done and CAPA is initiated
  • Personnel in the laboratory should test the effects if there is any malfunctioning and implement corrective and preventive action
  • The calibration data is checked as to how much product is tested or inspected?
  • If the equipment is suspected of producing results out of specifications then the product should be tested?
  • If the product is already shipped then supervisor should have a plan of action on how they recall the product? 
  • When the equipment failure is not in proportion to its function then such case of failure may be attributed by peripheral equipment
  • Before using the equipment again it should be tested again for producing correct results

Other considerations  

  • Even if it is an instrument or equipment proper procedures should be followed according to the SOP.
  • Records should be maintained along with the identification number of equipment or instrument.
  • Manufacturers name, identification type, serial number and unique identification number should be included in the report
  • Calibration should be in compliance with SOPs
  • Instructions given by the manufacturer
  • Dates and results of calibration or verification done to equipment or instrument
  • Historical data regarding their malfunction, modification or repair
  • Depending on the type of instruments or equipment’s they should be protected from fungal or microbial growth and should not be stored at extreme temperature or humidity
  • They should be protected to resist corrosion

Tags

Out of calibration equipments, instruments, root cause analysis, capa, calibration

References

www.ikev.org

www.ncdoj.gov

apps.who.int

View / Download