A procedure to operate and maintain the Laboratory Temperature OR Relative humidity monitoring System

To provide a procedure to operate and maintain the Laboratory Temperature /Relative humidity monitoring  System (Temp/RH Data monitor I.R.A ) furnace.

Sharing is caring, show love and share the thread with your friends.

Description

I.    PURPOSE :

To provide a procedure to operate and maintain the Laboratory Temperature /Relative humidity monitoring  System (Temp/RH Data monitor I.R.A ) furnace.

II.    SCOPE:
The procedure is applicable to I.R.A make Temp/RH Data monitor being used for monitoring of Temperature and Relative Humidity in the Quality Control laboratory

III.    RESPONSIBILITY:
        It is the responsibility of the Quality Control personnel to get the sensors  calibration done by a qualified external agency as per the schedule provided in this Standard Operating Procedure.

 IV.    DOCUMENT REFERENCE:
SOPs    :    Core procedure, Calibration of Equipment 
Forms    :    Temperature and relative humidity monitoring form.

 V.    OPERATION:
1.0      PRELIMINARY CHECK:

1.1        Check that the data is being transferred from the sensors to the data logger or not this
           is  checked by display of all the channels data in the screen of Data logger.

 2.0     BASIC OPERATION: 

2.1        Switch on the equipment continuously by connecting it to U.P.S.

2.2    The data is collected in 10 channels from different parts of quality control laboratory  
              described as below 

Channel number

Data type

Location in QCD

  1.  

Temperature

Instrumentation

  1.  

% Relative Humidity

Instrumentation

  1.  

Temperature

Wet analysis

  1.  

% Relative Humidity

Wet analysis

  1.  

Temperature

Gas chromatography

  1.  

% Relative Humidity

Gas chromatography

  1.  

Temperature

Non -routine

  1.  

% Relative Humidity

Non -routine

  1.  

Temperature

Spectroscopy

  1.  

% Relative Humidity

Spectroscopy

 

2.3    Take print out of data for every 2 hours on continuous basis.
2.4    Monitor the data daily once.
2.5    Any excursions beyond 200C to 300C and Relative humidity 30% to 70% R.H Should be noted and should be informed to maintenance department for corrective action .
2.6         The above excursions should be rectified with in more than 24 hours.
                                        
VI.   CALIBRATION:
        Calibration Frequency: Once in six months by external agency.
1)    The calibration should be carried out by external agency

 

Tags

Laboratory temperature, relative humidity, monitoring system, external agency