Calibration of Auto titrator

To lay down the procedure for Calibration of Auto Titrator.  

Sharing is caring, show love and share the thread with your friends.

Description

1.0    PURPOSE: 

1.1    To lay down the procedure for Calibration of Auto Titrator. 

2.0    SCOPE

2.1     This procedure is applicable for Calibration of   Auto Titrator in 

           quality control.

3.0    RESPONSIBILITY

3.1    Assistant-QC

4.0    ACCOUNTABILITY

4.1     Sr.Executive-QC

5.0    PROCEDURE

5.1    Operate the instrument as per the respective SOP and calibrate the Auto Titrator   as follows.

5.1.1     PRECISION

5.1.1.1 Determine the normality of 0.1 N NaOH/0.1N Perchloric acid 

solution  by taking  ~150 mg  of Potassium hydrogen phthalate. 

Calculate the % RSD five such determinations. It should be not 

more than 1.0. Record the details as per Format No.xxxxx

5.1.2    LINEARITY

5.1.2.1 Titrate 0.1 N NaOH/0.1N Perchloric acid solution with potassium 

    hydrogen phthalate by taking sample weights of 100, 125, 150,175

    and 200 mg. Plot a linearity graph of sample weight Vs volume 

    consumed. 

The correlation coefficient of the graph thus obtained shall  be NLT 0.999.Record the details as per Format No.xxxxxsop

5.1.3     FREQUENCY: Once in three months.                        


 

Tags

Auto titrator, perchloric acid, potassium, linearity graph,