1.0 PURPOSE:
1.1 To lay down the procedure for Calibration of Auto Titrator.
2.0 SCOPE
2.1 This procedure is applicable for Calibration of Auto Titrator in
quality control.
3.0 RESPONSIBILITY
3.1 Assistant-QC
4.0 ACCOUNTABILITY
4.1 Sr.Executive-QC
5.0 PROCEDURE
5.1 Operate the instrument as per the respective SOP and calibrate the Auto Titrator as follows.
5.1.1 PRECISION
5.1.1.1 Determine the normality of 0.1 N NaOH/0.1N Perchloric acid
solution by taking ~150 mg of Potassium hydrogen phthalate.
Calculate the % RSD five such determinations. It should be not
more than 1.0. Record the details as per Format No.xxxxx
5.1.2 LINEARITY
5.1.2.1 Titrate 0.1 N NaOH/0.1N Perchloric acid solution with potassium
hydrogen phthalate by taking sample weights of 100, 125, 150,175
and 200 mg. Plot a linearity graph of sample weight Vs volume
consumed.
The correlation coefficient of the graph thus obtained shall be NLT 0.999.Record the details as per Format No.xxxxxsop
5.1.3 FREQUENCY: Once in three months.
Auto titrator, perchloric acid, potassium, linearity graph,