OPERATION OF REVERSE LAMINAR FLOW UNIT

To have a written procedure for the usage of reverse laminar flow unit.

Sharing is caring, show love and share the thread with your friends.

Description

I.    PURPOSE :

        To have a written procedure for the usage of reverse laminar flow unit.

 II.    SCOPE:

        This procedure is applicable for the Micro-Kleanz make reverse laminar flow unit used in the Quality  Control department. 

III.    RESPONSIBILITY:

        It is the responsibility of Executive Quality Control to follow this SOP.
IV.    DOCUMENT REFERENCE :
             SOP     :  None
             Form      :  None

 V  .      PROCEDURE:
        Operating procedure
1.    Reverse laminar air flow unit works with theprinciple   of reverse laminar flow, which is  helpful to control the material exposure .
2.  Before starting the work  on reverse laminar flow ,switch on blow and observe pressure  obtaining  from magnahelic gauge.
3.    The pressure when the instrument is in operating condition should be between 5 to 15 PSIG.
4.    For any composite sample preparation of raw   materials, intermediates and finished product,should be Operate under the air flow only.

5.     Spread filter paper on the bench of reverse laminar flow while handling the material.
6.     Discard the exess materils and the filterpaper in the specified containers.
7.    After completion of the work, switch of the air blow and clean the surfaces using the  mop wetted in methanol/water 
       Maintenance:
1.    Clean the pre filters using water and air  monthly once or when ever the pressure is fallen below 5 mm of water
2.    Replace the HEPA filters for every three years

 

Tags

Laminar flow unit, material exposure, methanol, magnahelic gauge