img

Analytical method validation

    To provide a procedure for the validation of analytical methods under use 

Sharing is caring, show love and share the thread with your friends.

Description

I.    PURPOSE :

    To provide a procedure for the validation of analytical methods under use

II.    SCOPE :

The procedure is to guide the analytical method validation studies of all the analytical methods being developed and used for the analysis of different products. The procedure also provides the steps involved in the qualification of the compendial methods.

III .    RESPONSIBILITY :
A.    Quality Control shall :

i. Prepare, Review and approve the method validation protocol with an assigned protocol number.
 ii.  Carry out analysis & provide supporting data with a conclusive report on method validation carried out.
B.    Quality assurance shall :
i. Review and approve the method validation protocol.
ii. Should authorized any documents being used  during the validation study 

 ii. Review and approve the method validation report   

IV.DEFINITIONS :

 Method :  A set of all the written procedures and instructions involved in the  collection, processing, storage and analysis of chemical matrix for an analyte.

System Suitability :  Specific tests to ascertain the suitability and effectiveness of the operating system when employing chromatographic methods such as pressurized     liquid chromatography and gas chromatography. 

 Validation:  The obtaining and documenting of evidence to demonstrate that a method can be relied upon to produce the intended result under stated conditions within defined limits.  It is the process of establishing that the performance characteristics of a method (expressed in terms of analytical parameters) to meet the requirements for the intended application of the method. The normal analytical parameters used in method validation are defined as follows .

    Protocol:  A pre approved document ,which should define the scope, objective, responsibilities, method details , experimental design and acceptance criteria  with  consideration to the intended application of the study is being performed.
     
    Report:   An approved document ,which should explain the details found during the study and also should be concluded the status of the study against to the predetermined  acceptance criteria 

 Specificity:  The ability of a method to measure accurately and    specifically  the  analyte (the constituent being tested or analysed) in the presence of components that may be expected to be present in the sample.  It is a measure of the degree  of interference in the  analysis of complex sample mixtures such as blank, bulk drug substances containing 
 degradation products, related substances etc.    
Precision:  The degree of agreement among individual test results when a method is applied to multiple samplings of a homogeneous sample.  It is a measure of either the degree of reproducibility (agreement under different conditions) or of repeatability (agreement under same conditions) of the method.  

Limit of detection (LOD) :  The lowest concentration of analyte in a sample that can be detected, but not necessarily quantitated under the stated experimental conditions.  It is a parameter of limit tests.  
Limit of Quantitation (LOQ) ::  The lowest concentration of analyte in a sample that can be determined with acceptable precision and accuracy under the stated experimental conditions.  It is a parameter of quantitative determinations of low levels of compounds in sample mixtures, such as impurities in bulk drugs .

Linearity: The ability of a method to produce results that are directly or indirectly proportional to the concentration of the analyte in samples within a given range. 

Range :  The interval between the upper and lower levels of analyte (including these levels) that have been shown to be determined with precision, accuracy, and linearity using the method as written..
Accuracy:   The closeness of test results obtained by a method to the true value.  It is a measure of the exactness of the method.  

Ruggedness ( Intermediate precision ) : The Intermediate precision of an analytical method is the degree of reproducibility of test results obtained by the analysis of the same samples under a variety of normal test conditions ( with the combination of different analyst, different instrument, different lots of reagents, different day at normal laboratory conditions etc.)  Intermediate precision is normally expressed as the lack of influence on test results of operational and environmental variables of the analytical method.  

Robustness : The robustness of an analytical procedure is a measure of its capacity to retain unaffected by small, but deliberate variations in method parameters and provides an indication of its reliability during normal usage.

V.  PROCEDURE :

 01.    Before an analytical method is validated, the method must be completely developed  and a detailed procedure shall be documented.

 02.    A method must be validated for intended use, employing an approved protocol.

03.    Based on the analyte, the technique, a specific procedure shall be identified to assure the optimum operation of the system employed (as system suitability).

 04.    Validation studies shall be unique for each method, and shall be consistent with the methods purpose, concentration range to be tested, the sample and the technique used for analyte determination.

 05.    Validation experimental plans (protocols) shall be supported by appropriate statistical methodologies to assist in the interpretation of the results obtained.

06.    Any modification of an analytical method shall be subject to revalidation.

  07.    Documentation of the successful completion of laboratory studies is a basic  requirement for determining whether a method is suitable for its intended applications.

08.    The procedure used to evaluate each validation parameter as described in the approved protocol will be focused into the following steps :

i. Identification of appropriate validation parameters.

ii.    Design of experimental plan for parameters evaluation (validation protocol).

iii.    Execution of the validation.

iv.    Documentation to support of the conclusion that the method has been validated.

v.    The documentation shall contain the following :

a.    Approved  protocol

 b. Raw data – Analysis work books/test data sheet, Chromatograms,Charts, Thermogram, Histograms (if any) etc., 

 c.  Validation  report with conclusion of the study.

09.    The experimental design and acceptance criteria of validation parameters can be altered to other than the  conditions specified in this SOP

 VI. EXPERIMENTAL DESIGN:

A)    For Inhouse methods   :

1.1     Determine the  category (see Table I ) of the test/s which are to be validated.  Prepare a Method Validation Protocol in the specimen format given in the Annexure for each item.  Indicate in the protocol the experimental design comprising the tests or methods to be validated, the relevant analytical performance parameters to be checked, on the basis of the category given in the Table, the experimental details for carrying out these checks and the acceptance criteria for each parameter.  Guidelines for experimental designs for some commonly done tests and assays are  given below

     

 Category of tests and selection of validation parameters for In-house methods  

 

 

 

Category

of test

Specifi-

city

Precision

Linearity

& range

Accuracy

Intermediate precision

Robust-ness

LOQ

LOD

 

Assay by HPLC 

Yes

Yes

Yes

No

Yes

Yes**

No

No

 

Chromatographic purity  by HPLC

 

Yes

Yes

Yes

Yes

Yes

Yes**

Yes

Yes

 

OVI &Residual   solvent  test

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Yes

 

 

Chromatographic purity (By TLC)

 

Yes

Yes

No

No

Yes

No

No

Yes

 

 

Assay

Potentiometry

 

Yes

Yes

Yes

No

Yes

Yes

No

No

 

 

Particle size by laser diffraction

No

Yes

No

No

Yes

Yes

No

No

 

 

Rinse analysis in cleaning validation

 

Yes

Yes

Yes

Yes

Yes

No

Yes

Yes

 

 

Swab analysis in cleaning validation

 

Yes

Yes

Yes

Yes

Yes

No

Yes

Yes

 

** Carry out stability study for both mobile phase and test solution.

 

 

 

 

 

 

 

 

 

 

 

B)    For Pharmacopoeial methods   :  
1.2    Determine the  category (see Table I ) of the test/s which are to be validated.  Prepare a Method Validation Protocol in the specimen format given in the Annexure for each item.  Indicate in the protocol the experimental design comprising the tests or methods to be validated, the relevant analytical performance parameters to be checked, on the basis of the category given in the Table, the experimental details for carrying out these checks and the acceptance criteria for each parameter.  

1.3    Guidelines for experimental designs for some commonly done tests and assays are  given below

Category of tests and selection of analytical parameters for Pharmacopoeial methods   

Category

of test

Specifi-

city

Precision

Linearity

& range

Accuracy

Intermediate precision

Robust-ness

LOQ

LOD

 

Assay by HPLC 

Yes

Yes

No

No

No

No**

No

No

 

Chromatographic purity  by HPLC

 

Yes

Yes

No

Yes

No

No**

Yes

Yes

 

Chromatographic purity (By TLC)

 

Yes

Yes

No

No

No

No

No

Yes

 

Assay by

Potentiometry

 

Yes

Yes

Yes

No

Yes

No

No

No

 

** Carry out stability study for both mobile phase and test solution.

 

 

Experimental design for the method validation of Assay by HPLC

 

Analytical parameters

Experimental design

Acceptance criteria

 

System suitability

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

.

1. Establish the system suitability through out the validation followed by 2 procedures.

 

 

 

 

 

 

 

 

 

 

 

2.Procedure1: Establish the system suitability Whenever the chromatographic system changed or disturbed.

 

 

3. Procedure2: When there is no disturbance in  chromatographic system , establish the online system suitability between each parameter by injecting the standard solution(single time) and calculate the %RSD for the values obtained from initial values and each online standard.

 

 

 

 

 

 

 

 

 

As applicable (As per method details)

e.g. : For chromatographic methods

suitable parameters such as

theoretical plates,  tailing

factor  must be determined.

 

 

 

 

 

 

2. As applicable (As per method

details)

 

 

 

3. The cumulative RSD for the values obtained from initial (system suitability)values and the individual online standard solution should be less than 2.0%

 

 

 

Experimental design for an assay by HPLC method validation

 

Analytical parameters

Experimental design

Acceptance criteria

 

   Specificity

 

  1. Diluent, impurity interference

 

 

 

2.      Separate injection of individual impurities for identification and assay in case of chromatographic method.

 

 

3.      Compare the Relative retention times of

      each component obtained from individual

      solutions and  blend solution.

 

1.  There should be no interference of the diluent or impurities & the  RT of  drug substance.

 

2.The impurities of impurities should

   not  interfere at the retention time of   

   main compound

 

 

3. The relative retention times (with respect to drug substance) of each impurity obtained from the  blend solution should be comparable(within + 0.06) to the relative retention times obtained from the individual solutions of each impurity.

Precision

 

  1.  Establish the system suitability based on the condition

 

  1. Repeatability of 5 replicate determination of standard solution at assay level concentration.

 

 

3.  Method precision of atleast 6 replicate assay 

  dermination of a sample. Determine the      

  %RSD of assays.

1 System suitability as applicable

 

2   The % RSD for peak area response

     of replicate standards should not be

     more than 2.0 %

 

3   % RSD of assay should not be

     more than 2.0 %

 

 

Experimental design for the method validation of Assay by HPLC

 

Analytical parameters

Experimental design

Acceptance criteria

Linearity & range

  1. Determination of minimum 7 levels of standard in the range of at least 50 to 150% (50%,80%,90%,100%,110% 120% and 150%) of the assay level concentration .Carry out linearity for first & last level in five replicates & other levels in triplicate and take mean response for calculation and also calculate the % RSD for peak area response of three replicates of each level .

 

 

 

  1. Determine linearity co-efficient and % y intercept.

 

 

 

 

 

  1. Determine the precision of lower (1st level) and higher level (last level) in five replicates.
  1. 1.  Linearity co-efficient of mean response of replicate determination plotted against respective concentration should not be less than 0.999.
  2.  
  3. The %y-intercept as obtained from the linearity data should be within ± 2.0 % determined using the formula :
  4.  

           100 x y intercept

   =   -------------------------------------

           Response of assay level

 

 

 

2. The % RSD for peak area response

    of three replicates of each level 

     should not be more than 2.0 % .

 

 

 

 

3.  The % RSD for peak area responses

     of five replicates of first & last level

     should not be more than 2.0 %

 

 

                       Experimental design for the method validation of Assay by HPLC

 

Analytical parameters

Experimental design

Acceptance criteria

 

Ruggedness (Intermediate 

 precision)

 

  1. Establish the system suitability

 

 

 

 

 

  1.  Repeatability of 5 replicate determination of 

Standard solution at assay level concentration.

       Determine the % RSD of peak area response

       & retention time.                                 

 

 

 3.  Intermediate precision of Six replicate 

    determination of assay of a   sample (Which is

    assayed  under precision study) with the

    combination of different  analyst / day  /

    instrument / different column of same make. 

    Determine the  %RSD of assays.

 

 

 

4.   Determine the cumulative% RSD   of assays of 

     precision study and Intermediate precision study

     together. 

 

1.System suitability as

  applicable

 

 

2 . The % RSD for peak area

response of replicate standards should not be  more than 2.0 % 

 

 

 

3 % RSD of assay should not be

     more than 2.0 %

 

 

 

 

 

 

 

4 Cumulative % RSD of assay of precision study &intermediate

 precision should not be more than 

  2.0 %

 

 

Experimental design for the method validation of Assay by HPLC

 

Analytical parameters

Experimental design

Acceptance criteria

Robustness

 

  1. Deliberately alter any 2 critical parameters(as per requirement) by  minor variation.

 

a      Mobile phase composition (higher

        concentration by 2 % or lower

        concentration by 5-10 %)

 

b      pH of mobile phase (± 0.2 )

 

c     Column oven temperature (±  2.0 °C)

 

d      Mobile phase flow rate   

       

  1. Establish the system suitability   

 

 

3. Method precision of three replicate  determination    of assay of a   sample (Which is analysed under precision  study) under each individual condition of the robustness study . Determine the  %RSD of assays.

 

 4.Determine the cumulative % RSD of Assay of  Precision study & Robustness study together.

 

1.  Select as per the applicability of the

    method

 

 

 

 

 

 

 

 

2.   System suitability as applicable.

    ( The % RSD for peak area response

     of replicate standards  should not be

     more than 2.0 % 

 

3   % RSD of assay should not be

     more than 2.0 %

 

 

 

4.  The cumulative % RSD of assay of

     precision study along with the

     results of Robustness study should

     not be more than 2.0 %

 

 

Experimental design for the method validation of Assay by HPLC

 

Analytical parameters

Experimental design

Acceptance criteria

Solution stability

 

  1. Establish the system suitability   

 

  1. Prepare two samples of a batch and

      calculate the assay of each preparation

     and note down the result as initial assay

      and store one sample is on bench  top and

      another one is in refrigerator simultaneously

      for about 24 hours. After the specified time

      interval evaluate the assay of each sample

      using the same chromatographic 

      conditions(same mobile phase, column

      &system) with freshly prepared standard

      solution.   Also monitor the  pattern of 

       chromatogram of sample solution 

    

 

  1. When there is no solution stability for 24 hours in  any of the above said conditions, determine the solution stability at predetermined time interval ( 2 hours, 4 hours, 8, hours,12 hours ) at laboratory temperature. If sample is sensitive carry out solution stability study by keeping the sample solution in refrigerator at 2-8 °C at same time interval of laboratory temperature. Note: After finding the time interval of solution at which is not stable no need to evaluate solution stability for the samples of remaining time intervals.

The time intervals can be reduced and altered for sensitive drug substances .

 

 

 

 

 

 

 

 

1   System suitability as applicable.

  

 

2. a. The % variation between the initial

    results and the results of solution

    stability(individually) should be less

    than 1.0%

 

   b .If any extraneous peak of impurities

    or degradants are observed in     

    chromatogram of sample than that of

    initial pattern  consider the solution

    is not stable. 

 

 

 

 3. Report the solution stability in hours

(solution stability is the last time interval of the test solution is found stable)

 

 

 

 

 

 

 Experimental design for the method validation of Assay by HPLC

 

Analytical parameters

Experimental design

Acceptance criteria

Mobilephase stability

 

  1. Prepare  mobile phase and store one portion in refrigerator and the other portion is at laboratory temperature for  at least two days

 

  1. Establish the system suitability

 

  1.  Repeatability of 5 replicate determination  

       of Standard solution at assay level

       concentration.

 

4.    Method precision of three replicate     

      determination of assay of a sample

     (Which is assayed  under precision study)

       using each storage condition of  mobile

       phase.  Determine the  %RSD of assays.

 

 

5.   Determine the cumulative% RSD of

     assays of each condition of mobile phase

     stability study and  precision together. 

 

1. Establish the mobile phase stability

    for 2 to 7 days  at each condition

 

 

 

 2. System suitability as applicable

 

3 .The % RSD for peak area response

of replicate standards should not be more than 2.0 % 

 

4.  % RSD of assay should not be

     more than 2.0 %

 

 

 

 

 

5.   Cumulative % RSD of assay of

     precision study & each condition of

     the mobile phase stability study

     should not be more than 2.0 %

 

 

  Experimental design for the method validation of Assay Titremetry  

Analytical parameters

Experimental design

Acceptance criteria

Specificity

Check the blank interference of  diluent.

 

.

 

The Burette reading value should be negligible ( Around 0.1 ml ).

* Blank reading should be taken for the   

   calculation of assay .

Precision

Method precision  of at least 6 replicate assay  determination of a sample. Determine the % RSD of assays.

 

   % RSD of assay should not be

     more than 2.0 %

 

Linearity & range

 

Determination of minimum 6 levels of standard in the range of at least 50 to 150% of the assay level concentration including assay level concentration. Carry out linearity for first & last level in six replicates & other levels in triplicate and take mean burette reading  for calculation. Determine linearity co-efficient and % y intercept.

 

 

Determine the precision of assay content  of lower (1st level) and higher level (last level) in six replicates.

Linearity co-efficient of mean burette reading  of replicate determination plotted against respective concentration should not be less than 0.999.The %y-intercept as obtained from the linearity data (without extrapolation through origin 0,0) should be within ± 2.0 % determined using the formula :                 100 x y intercept

   =   -------------------------------------

           B.R. of assay level

    The % RSD for % assay  of

     replicate (1st level)   and higher level 

    (last level) should not be more than 2.0 %.

 

Ruggedness (Intermediate 

 precision)

Reproducibility   of Six replicate determination of assay of a    sample (Which is assayed under precision study) by a different analyst / day at normal laboratory condition / instrument (if available)Determine   the cumulative % RSD   of assays of precision  study and Intermediate precision study together. 

1   System suitability as per method

 

2   % RSD of assay should not be

     more than 2.0 %

 

3   Cumulative % RSD of assay of

     precision study & intermediate

     precision should not be more than 

     2.0 %

 

Experimental design for the method validation of chromatographic purity/

                  Related substances by HPLC

Analytical parameters

Experimental design

Acceptance criteria

 

System suitability

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

  .

1. Establish the system suitability through out the validation followed by 2 procedures.

 

 

 

 

 

 

 

 

 

 

2.Procedure1:Establish the system suitability when ever the chromatographic system changed or  disturbed.

 

 

3. Procedure2: When there is no disturbance in  chromatographic system , establish the online system suitability between each parameter and observe the compliance of system suitability.

 

 

 

 

 

 

 

 

 

1. As applicable (As per method details)

  e.g. : For chromatographic methods 

    suitable parameters such as  

     resolution, RRT , Peak to valley ratio

    must be determined. If not specified keep

    following criteria .

 

 

   

 

 

 

 

2. As applicable (As per method

    details)

 

 

 

3. Should complies the system suitability  

 

Experimental design for the method validation of chromatographic purity/

        Related substances by HPLC

Analytical parameters

Experimental design

Acceptance criteria

 

Specificity

 

 

 

 

 

 

<

Tags

Analytical method validation, validation

View / Download