To lay down the procedure to be followed for the operation, calibration and maintenance of Air sampler.

Sharing is caring, show love and share the thread with your friends.

Description

 

  1. Purpose    :                    

To lay down the procedure to be followed for the operation, calibration and maintenance of Air sampler.

  1. Scope        :

This procedure applies for following Air samplers:

Make

Model

Himedia

LA053

Merck

MAS-100 NT         

 

 

 

 

 

 

  1. Responsibility      :

The responsibility of operation and calibration of Himedia Air sampler and MAS-100 NT Air sampler lies with Microbiologist / Analyst.

  1. Definitions / Abbreviations:
  2.  Media: A substance used to provide nutrients for the growth and multiplication of microbes.
  3. Procedure:
  4. Operating Procedure for Himedia Air Sampler:
  5. Leave the air sampler under charge for about 14 hours before work starts.
  6. Set the timer for 10 minutes by using the following steps.
  7. To set the 1st digit, Press ‘SET’ in the remote, the 1st digit in digital display will blink as “0”00
  8. Press ‘B2’ in the remote twice to set the 1st digit as  “2”00.
  9. Press ‘B1’ in remote to set the 2nd digit, it will blink as  2 “0”0.  
  10. Press ‘B2’ again to set “0” in the 2nd digit.
  11. Press ‘B1’ again to set the 3rd digit, it will blink as 20 “0”.  
  12. Press ‘B2’ again to set “0” in the 3rd digit.
  13. Finally press ‘SET’ to obtain the setting time in the digital display 10 minutes. 
  14. After setting 10 minutes, place the SS cone inside the sampler.
  15. Fix the media containing petri plate without lid in SS cone.
  16. Finally fix the SS feeder cone, circular clamp assembly in position.
  17. Mount the sampler either horizontal/vertical position by using adopter to the MS tall stand.
  18. Press ‘START’ on the remote to start air sampling and the digital display starts showing“0” to “10 minutes” as the sampling period proceeds to set time.
  19. The air sampler stops to take an air when the digital display shows “10 minutes”.
  20. After completion of the air sampling, remove agar plate and close with lid.
  21. Repeat the same for other locations.
  22. Operating Procedure of Merck (MAS-100NT) Air Sampler:
  23. Before use of Air sampler MAS-100 NT is charged approximately 4 hours. Make sure that red point of the charging plug matches the black point of the handle.
  24. To start the instrument push the left down button for about 2 seconds the blue diode will light up.
  25. There is information on display panel as MAS-100 NT.
  26. After that on second display shows instrument information such as Name, Department    location, S.No. S.Header-300x0.6, Next calibration:12 months and there is option ‘OK’ press   it or after 3 seconds it shall automatically select OK option.
  27. On the third display shows current date and time with information of:

                         Mode: STD

.                       Volume: 1000 L

Start                Menu

  1. If  air volume program is pre-setted skip step 5.2.7 and follow the step 5.2.8
  2. Setting procedure for air volume in MERCK (MAS-100NT) Air Sampler:
  3. To set the required volume go to option ‘Menu’ press it there is list of options as below           

                           Shut down

                           Mode selection

                           Template data

                           Process setting

                           Standard volume

                           Delay

                           User selection

                           Print setting

                           Sample log

                           Alarm log

                           Config Device

                           System info

 

  1. Select the standard volume option with the help of  ‘?’ascending or ‘?’ descending buttons   on right side of the display panel.
  2. There is display of standard volumes as below

                           Volume 1: xxx

                           Volume 2: xxx

                           Volume 3: xxx

                           Volume 4: xxx

                           Volume 5: xxx

                                                            ↓

                           Select Volume 1

                          ↓

                           Press option ‘Change’

                                                                        ↓

                           Set the required volume (1000 L) by pressing ‘?’ascending or ‘?’ descending buttons

                                                                        ↓

                           Press option ‘Save’

 

  1. Remove the sampling header by rotating it in anticlock wise direction then there shall be dish (Plate) supporter with 3 blue jaws which shall be adjustable as per plate size by rotating the jaws screws with the help of ELENKEY.
  2. Place the prepared agar plate on the dish supporter and replace the sampling header by   rotating in clockwise direction.
  3. Remove the cover lid.
  4. Start the instrument by pressing button ‘Start’ on left side of the display screen.
  5. Instrument starts and there is display of ‘Measurement’  
  6.  Volume    : xxx

                                                                                                                                                                                                       Time Rem: xxx

                                                                                                                                                                                               State        : Running

        and blue diode shall be blinking up to the set volume reached.

  1. Remove the sampling header and replace with other agar plate and do air sampling.
  2. To shut down the instrument press option
  3.           ‘Menu’

                                                                                                                                                                                                            ↓

                                                                                                                                                                                                ‘Shut down’

 

  1. Incubate the plates as per procedure “Environmental monitoring for bioload in pharma and microbiology testing area” (Doc. Number:SOPxxxxx)

5.4    Record the results by considering the positive hole conversion table as given below.

Pr = N [1/N + 1/N-1 + 1/N-2 + 1/n-r+1]

Pr = Potable statistical total                N = Number of holes

r = Number of colony forming units counted on 90mm petri dish.

r

Pr

r

Pr

r

Pr

r

Pr

r

Pr

r

Pr

1

1

51

56

101

123

151

209

201

332

251

541

2

2

52

57

102

124

152

211

202

335

252

547

3

3

53

58

103

125

153

213

203

338

253

553

4

4

54

59

104

127

154

216

204

341

254

560

5

5

55

61

105

129

155

218

205

344

255

566

6

6

56

62

106

131

156

220

206

347

256

573

7

7

57

63

107

132

157

222

207

350

257

580

8

8

58

64

108

134

158

224

208

353

258

587

9

9

59

66

109

135

159

226

209

357

259

594

10

10

60

67

110

137

160

228

210

360

260

601

11

11

61

68

111

138

161

230

211

363

261

609

12

12

62

69

112

140

162

232

212

367

262

616

13

13

63

71

113

142

163

235

213

370

263

624

14

14

64

72

114

143

164

237

214

374

264

632

15

15

65

73

115

145

165

239

215

377

265

641

16

16

66

74

116

146

166

241

216

381

266

649

17

17

67

76

117

148

167

243

217

384

267

658

18

19

68

77

118

150

168

246

218

388

268

667

19

20

69

78

119

151

169

248

219

391

269

677

20

21

70

80

120

153

170

250

220

395

270

686

21

22

71

81

121

155

171

253

221

399

271

696

22

23

72

82

122

156

172

255

222

43

272

707

23

24

73

83

123

158

173

257

223

407

273

717

24

25

74

85

124

160

174

260

224

410

274

728

25

26

75

86

125

161

175

262

225

414

275

740

26

27

76

87

126

163

176

264

226

418

276

752

27

28

77

89

127

165

177

267

227

422

277

765

28

29

78

90

128

167

178

269

228

427

278

778

29

30

79

92

129

168

179

272

229

431

279

791

30

32

80

93

130

170

180

274

230

435

280

805

31

33

81

94

131

172

181

277

231

439

281

820

32

34

82

96

132

174

182

279

232

444

282

836

33

35

83

97

133

175

183

282

233

448

283

853

34

36

84

98

134

177

184

284

234

452

284

871

35

37

85

100

135

179

185

287

235

457

285

889

36

38

86

101

136

181

186

289

236

462

286

909

37

39

87

103

137

183

187

292

237

466

287

931

38

41

88

104

138

184

188

295

238

471

288

954

39

42

89

105

139

186

189

297

239

476

289

979

40

43

90

107

140

188

190

300

240

481

290

1006

41

44

91

108

141

190

191

303

241

486

291

1036

42

45

92

110

142

192

192

306

242

491

292

1069

43

46

93

111

143

194

193

308

243

496

293

1107

44

47

94

113

144

196

194

311

244

501

294

1150

45

49

95

114

145

198

195

314

245

507

295

1200

46

50

96

115

146

200

196

317

246

512

296

1260

47

51

97

117

147

202

197

320

247

518

297

1335

48

52

98

118

148

203

198

323

248

523

298

1435

49

53

99

120

149

205

199

326

249

529

299

1585

50

55

100

121

150

207

200

329

250

535

300

1885

                         

 

Calibration:

  1. Air samplers shall be calibrated by external agency.
  2. Frequency of calibration: Once in a year ± 30 days.
  3. Maintenance:
  4. The external and internal surface of the air samplers shall be cleaned with 70% IPA.
  5. Perforated lid and dust cover of MAS-100NT and SS cone feeder of Himedia air sampler shall be autoclaved for 30 minutes at 121ºc.
  6. In case of breakdown call the service engineer for rectification.
  7. Related Documents:

Environmental Monitoring for Bioload in Pharma and Microbiology Testing area (Doc. Number: SOPxxxxx

 

Tags

Microbiology, procedure, operation, air sampler

References

View / Download