img

Operation and Calibration of UV Cabinet Advance

To provide a procedure to operate and maintain the Advance make UV Cabinet. 

Sharing is caring, show love and share the thread with your friends.

Description

I.    PURPOSE :
     To provide a procedure to operate and maintain the Advance make UV Cabinet.

II.   SCOPE :
       This procedure is applicable for the UV cabinate used for TLC testing in the Quality Control department 

III.    RESPONSIBILITY :
    It is the responsibility of all the personnel involved in the analysis using of the equipment.
    
IV.    DOCUMENT REFERENCE :
    SOPs    :    Core procedure: Calibration of Equipment
    Forms    :    None

V.    OPERATION :           
1.0   PRELIMINARY CHECK :

1.1    Check that the instrument and surroundings are clean, if not, clean with a soft cloth duster.  
2.0 BASIC OPERATION : 

2.1    Before using  ensure the validity of calibration.
2.2    Switch  on  the  power.
2.3    Switch on the normal light and open the front door pulling towords the front side to keep the TLC plates
2.4     Place the TLC plate in UV chamber horizontally on the base of the chamber. 
2.5     Switch on the particular nm wave length light i.e.  254 nm or 365 nm light at which the TLC is to be viewed .
2.6     Do not expose to UV light while handling the UV cabinet, use always normal light only while operations.
2.7       Switch of the power after the completion of the test.
 
 VI. CALIBRATION  :                        

Calibration Frequency: First week of every monthCalibration of 254 nm lamp.1.    Weigh about 40 mg of sodium salicylate and transfer into 100 ml volumetric flask dissolve   and dilute to  100 ml with methanol.
2.   Apply 5 ?L of the above solution on a TLC plate (glass or aluminum) with adiameter of about 5 mm.


3.    Examine the spot in UV cabinet under 254 nm keeping the plate normal to the radiation.
4.    The standard spot should be revealed without doubt i.e clearly.Calibration of 365 nm lamp.
1.    Weigh about 200 mg of sodium salicylate transfer into 100 ml volumetric flask dissolve and  dilute to 100 ml with methanol.
2.    Apply 5 ?L of the above solution on a TLC plate (glass or aluminum) with a diameter of about 5 mm.
3.    Examine the spot in UV cabinet under 365/366  nm keeping the plate normal to the  
       radiation.
4.    The standard spot should be revealed without doubt i.e clearly.
5.    If the instrument fails in calibration, proceed as per the Core Procedure, “Calibration of Equipment”.

Acceptance criteria : 

01. 0.2% spot should be visible at 365nm and 0.04% spot should be visible at 254nm .

Tags

uv cabinet, salicylate, core procedure, salicylate

References

View / Download