img

Handling of HPLC columns

This document describes the procedure for purchase, installation, storage, performance evaluation, usage, cleaning and  withdrawl of HPLC columns.

Sharing is caring, show love and share the thread with your friends.

Description

PURPOSE :
This document describes the procedure for purchase, installation, storage, performance evaluation, usage, cleaning and  withdrawl of HPLC columns.

    II.    SCOPE :
Applicable to all HPLC columns used in the Quality Control department. 

III.    RESPONSIBILITY :
It is the responsibility of the Quality Control personnel involved in the usage of HPLC columns, to follow the procedure as mentioned in this Standard Operating Procedure.
IV.    DOCUMENT REFERENCE :
         SOPs    :   None
         Forms    :   HPLC Column inwards register, Column performance work book & tag.

V.    PROCEDURE :


1.0      PURCHASE OF COLUMNS :                                       Responsibility  : Department Head

1.1         Indent & receive the columns from Purchase department and verify whether correct columns are received against to Specify the column make, brand name, dimensions and particle size.

1.2        Check the column for physical damage which may have occurred during shipping.

2.0   INSTALLATION OF COLUMNS               Responsibility  : Section Head
        
2.1        Retain safely the column performance certificate sent by the manufacturer.

2.2    Ensure that the column ends are securely closed.

2.3    Store the column in the designated area.

2.4    Test the column immediately to verify quality  and allot individual 

3.0      COLUMN PERFORMANCE EVALUATION    Responsibility  : Section Head/Analyst

3.1    Prior to column performance, flush the column with suitable solvent (compatible) based on the solvent in which the column is shipped. 
3.2     Read and follow the instructions such as temperature, pH, back  pressure limits mentioned for one particular column.
3.3       Test the column for performance by the following procedure :

3.3.1     For reverse phase bonded packings (C18, C8, Cyano, Amino etc.)
3.3.1.1     Uracil stock solution : Weigh accurately about 50 mg of Uracil in a    50 ml volumetric flask.  Dissolve in 25 ml mobile phase and dilute upto the mark with the mobile phase.

3.3.1.2    Standard solution :  Weigh accurately 40 mg of phenol and 60 mg of anisole transfer     
                  in a 50 ml volumetric flask, containing about  20 ml of mobile phase and dissolve.  
                  Add 2 ml of uracil stock solution and dilute up to the mark.
3.3.1.3      Chromatographic conditions :
                Mobile phase :  Methanol : water (70 : 30)
                Flow : 1.0 ml/minute
                Wavelength : 254 nm
                Injection volume : 20 µl

3.3.1.4    Procedure :  Inject the standard solution 5 times.  Calculate the column efficiency (theoretical plates). Symmetry factor and capacity factor for the peaks of phenol and anisole.  Use the retention time of uracil as void volume for determination of capacity factor.  Relative standard deviation of peak area for 5 replicate injections should not be more than 2.0% for phenol and anisole.

 3.3.1.5     Acceptance criteria :   Capacity factor      :  For phenol : Not less than 0.30
                                                                                       For Anisole : Not less than 1.00
                                                 Theoretical plates  :    For phenol : Not less than  2000
                                                                                       For Anisole : Not less than 2000
                                                              USP tailing : Not more than 2.0 (each for phenol and anisole)

3.3.2      For Adsorption/Normal Phase columns

3.3.2.1    Preparation of the standard :  Weigh about 200 mg each of acetophenone ,80 mg of 2,4-dinitrotoluene and 200 mg of toluene in a 100 ml volumetric flask.  Dissolve in 25 ml of mobile phase and dilute upto the mark with mobile phase.  Dilute 3 mL of the above to 100 mL in a volumetric flask with mobile phase.

3.3.2.2     Chromatographic conditions
              Mobile phase : Hexane : Ethyl acetate (98:2)
              Flow : 1.0 ml/minute
              Wavelength : 254 nm
              Injection volume : 20 µl
              Run time              : 20 Minutes for 250mm and above 250mm length columns. 
                                           10 Minutes for below 250mm length columns.

3.3.2.3    Procedure :  Inject the standard solution 5 times.  Calculate the column efficiency (theoretical plates), symmetry factor and capacity factor for the peaks of acetophenone and 2,6-dinitrotoluene.  Use the retention time of toluene as void volume for determination of capacity factor.  Relative standard deviation of peak area for 5 replicate injections should not be more than 2.0% for acetophenone and 2,4-dinitrotoluene.
 3.3.2.4   Acceptance criteria :  
                    For Below 250 mm length columns
               Capacity factor : For acetophenone : Not less than 0.15
                                   For 2,4 dinitrotoluene : Not less than 0.30

                      Theoretical plates  : Not less than 1000 (Each for  acetophenone and 2,4 dinitrotoluene)

                                 USP tailing : Not more than 2.0 (Each for acetophenone and 2,4 dinitrotoluene)

               For  250 mm and above 250mm length columns
            Capacity factor : For acetophenone : Not less than 0.40
                                For 2,4 dinitrotoluene : Not less than 2.0

                    Theoretical plates  : Not less than 1000 (Each for  acetophenone and 2,4 dinitrotoluene)

                    USP tailing : Not more than 2.0 (Each for acetophenone and 2,4 dinitrotoluene).

3.4      After completion of the performance test, attach a tag to the column indicating the column number, date of  performance check      
           and signature.

3.5    Update the index of column to include the new column.

3.6    If the column fails to meet the acceptance criteria, then return the column to the supplier and arrange for replacement. 

3.7    After meeting the requirements of column performance allot the ID no as given below

3.8      QCD/LC/xxx  ,  xxx is the serial number.
3.9      Dedicate the columns individually for finished product and  stability testing of each 
           product (where ever possible) and label the same on the label.
4.0        USAGE OF COLUMNS                                           Responsibility  : Analyst

4.1    Handle the column with extreme care.  Never allow the column to experience mechanical shocks or vibrations.
4.2    Select the column by referring to the HPLC column list and as indicated in the specification.
4.3    Allow the column to stabilize on the system under the set parameters of the instrument and the mobile phase.

4.4     Perform the analysis on the column as recommended in the ‘Tests and  Methods’ given under specification.

4.5     On completion of analysis, flush and clean the column with the relevant solvents as recommended in the section on cleaning and care of HPLC columns.

4.6         Replace back the column to the designated place .

     5.0    CLEANING OF COLUMNS                            Responsibility  : Analyst

 5.1    Always use filtered and degassed mobile phase/ solvents.

5.2    Before change over of a mobile phase, make sure that solvents used for column flushing  are miscible with the mobile phase.

5.3    Equilibrate the column for about 30 minutes at the flow rate as specified in the individual specification.  If ion-pairing agents or additives are used, flush the column for about 1 hour at the specified flow rate for complete equilibration.
5.4    After completion of analysis, flush the column with the appropriate solvents (as specified in the individual column log).  If the mobile phase contains buffer, flush the column first with sufficient quantity of water and then with an organic solvent, used for mobile phase preparation.

5.5    For longer storage, store the column in 100% methanol and be sure that the end plugs are firmly in place.  Never let the column dry out.

5.6     Storage conditions for HPLC columns :

Column type

Storage solvent

Reverse phase

(C18, C8, C4, C2, C1, Phenyl)

100% methanol

 

Normal phase

(Silica, CN, NH2, Diol, Alumina)

Isopropanol or n-hexane

Ion exchange

(SAX, SCX)

Methanol

(Flush the column with 50 ml water prior to storage solvent)

Size exclusion

(Diol)

0.05% sodium azide in water or in 10% methanol

 

While changing the mobile phase, gradually increase the flow rate of the new mobile  phase in increments from 0.1 ml/minute upwards.
5.8         A shift in retention or resolution is an indication of insufficient flushing of the column.  Flush with a organic solvent which may be a part of the mobile phase of the product for which the column is dedicated.  
5.9        If  buffers like citrate or ammonium carbonate are used in the mobile phase, ensure that the HPLC columns which can withstand the conditions of analysis or the specified HPLC columns as given in the specification are used.  Other HPLC columns may get deactivated in such conditions (condition this point while selecting a column for particular applications).
 6.0    WITH DRAWAL OF COLUMNS       Responsibility  : Department Head/Section Head

6.1    Withdraw the column from usage in case of any of the following instances:

6.1.1    Physical damage 

6.1.2    Failure to comply system suitability requirements as per specification .( Note: When failure to the system suitability requirements as per specification, evaluate once again the column performance and take the decision of withdrawl )

6.1.3     Failure to comply with performance check requirements.

6.1.4    Indicate the withdrawal on the column inwards register and update the column index.

Tags

Hplc columns, purchase, installation, performance evaluation, suitability

References

View / Download